COMPARISONS
|
SETTINGS |
PHILLIPS 5500 |
GE Vivid 7 |
ACUSON Sequoia512 |
|
FREQUENCY |
Harnonics-range F1-send-1.3/receive-3.2-more P F2-send-1.6/3.2-P/T F3-send 1.3/2.6-T F4-send 1.6/3.2-T.R F5-send 1.6/3.2-R |
Harmonics range is send
1.5/receive 3.0 Send 1.7/receive 3.5 Send 1.9/receive 3.8 Send 2.0/receive 4.0 |
Harmonics range is send
/receive 2.25 /receive
2.75 / receive 3.25 /receive
3.75 /receive
4.25 |
|
IMAGE DEPTH |
24cm is the max. depth |
30cm |
30cm |
|
Gray scale levels |
14 steps-post proc. A-N |
18 levels-Dynamic range is
20 to 2 |
16 steps-post proc range
is 0 to 7 |
|
Automatic tissue Optimization available |
No |
Yes-called Tissue Optimization-press down on
2D gain |
Yes-called TEQ-press
button under sliding gain levers |
|
Lateral gain available |
Yes |
No |
No |
|
Edge enhancement available |
No |
Yes-called contour or edge
enhance |
Yes-use edge toggle Or dial to Valve preset |
|
Probe |
S3-Phased array |
M3S/ M4S-matrix array |
4VIC-s-vector wide view
array |
|
Respiratory waveform
available |
Yes, go under physio |
Yes-go under physio |
Yes-only with a nasal
respirometer |
|
Post-processing Available for additional
measurements |
Yes—under loop and disk
retrieve-can do 2D, Doppler, and colorize. Can not add
color, m-mode or change sweep speed |
Yes-go under review and
you can do 2D, M-MODE, TVI , Doppler-can
change sweep speed and color baseline, can add B color maps---but can
not add color |
Yes-go under review and
you can do 2D, Doppler. Can not
add color, m-mode, colorize and can not change sweep speed |
|
Extra features |
LIMA-on
stress machine 3D-on TEE machine |
TVI TSI 4D-on new machine |
VVI-on flat panel
machine |